Date | Component | Description | Status |
---|---|---|---|
20210922 | Focus lock | z focus is unstable, z jumps on the order of 50-200 nm. likely the laser | Solved by reducing the focus-lock laser power. Recommended to use 5 - 20 mW. |
20210922 | EMU - Micromanager 2.0 - z-stacks | z stacks do not acquire the correct amount of slices, but 2-3 slices less | validated on 28 oct 21 (markus) on git commit 0da620a4e5c472fa0b8edbfae3d71e8a44c07d9a (july 2 2021) - z-stack from -1.5 to 1.5 um with 0.01um spacing acquired 299 instead of 301 slices |
20210922 | EMU - Micromanager 2.0 - z-stacks | no laser illumination during z stacks. This is due to wrong synch signal on every first frame from OrcaFlash when ROI is cropped. | Use the following exposure, 25 ms, 50 ms, 100 ms, and so on, or trigger laser "ON" /mm. |
20210922 | Micromanager 2.0 - acquisition very laggy | live display of frames only works when cursor is positioned on top of the window, and even laggy then | new PC quote requested, might not be resolvable though |
20211118 | Micromanager 2.0 - acquisition unlocks focus randomly | randomly unlocks the focus in the middle of the experiment, and then sometimes, but not always, re-locks at a random point in time | |
20210922 | EMU - activation | After starting the experiment and activating the tick box for the automatic UV pulse length adjustment, the UV duration jumps to the maximum (or to the end value from the previous experiment?). This cannot be solved by manually dragging the slider down/setting the value to 0 before or after the experiment. | |
20210922 | Matlab - fitting impossible | Sometimes, the fitting fails with a CUDA error (device not available). In some cases, reducing the number of ROIs helps, but usually a full restart of the PC is required. | new PC quote requested from purchase |
20210923 | Matlab - Registration of two cameras for multi-color | Registrations between two cameras seem to change over time so the ratio calculations for color assignment gives poor separations. Effect of system warm-up seems not causing this, so could be due to z-position jumps. | Needs to be re-checked after focus lock becomes stable. |
20210923 | Automatic data backup to tier2 | Automatic data backup to tier2 is not working, and so needs to be done manually. Run the script on the Desktop manually. | There was an error in backup script. Fixed. |
Date | Component | Description | Status |
---|---|---|---|
20210922 | EMU - freezing | Blinking movie window freezes after ca. 40 000 frames. For longer acquisition also the EMU Gui freezes (not responsive, eventually all windows go white). It becomes almost impossible to then stop an acquisition, very laggy response. Workaround for now is to keep the acquisition running and hope that the activation in the background still performs correctly. | Noone working on it for now |
20210922 | Focus lock? - z-jumps | Philipp observed some z-jumps similar to what happens on M4. | Checking older files atm to see when it started, I think noone else works on it atm |
20210922 | Micromanager 2.0? - BFP lens | BFP sometimes gets stuck in beam path. Workaround is to just put it back in and out again. | Solved by replacing the BFP lens mount with a slider. |
20210922 | Micromanager 2.0 - Error hardware configuration | Everytime MM is started, it throws this error: "Failed to load hardware configuration. Error in device "COM9": Unknown error in the device (1)". One then just reloads the configuration and it works. Not too bad, just annoying. | Noone working on it for now |
15/02/24
Laser | Before fiber (mW) | After fiber (mW) | Coupling % |
---|---|---|---|
405 | |||
488 | |||
561 | |||
640 | |||
Booster |
02/06/20 (european "normal" expression: d/m/y)
Laser | Before fiber (mW) | After fiber (mW) | Coupling % |
---|---|---|---|
405 | 7.63 | 5.63 | 73% |
488 | 168 | 143.5 | 85% |
561 | 216 | 187 | 86% |
640 | 173 | 157 | 90% |
Booster | 133 | 117 | 87% |
Measurements before fiber were done right before the lens and after the polarization beam splitter.
If you get this message, click on ok and in the Micro-Manager menu "Configuration", load a new configuration: the "booster to normal" configuration. After loading, close Micro-Manager. Start the Toptica software (the red dot), connect to the Booster, click on reset clip and quit the software. Start Micro-Manager 2 again and load the normal M2 configuration.
Micro-Manager 2 often gives COMport errors on M2. Just try to "reload the hardware configuration" from the MM2 Configuration menu. In most cases it works! If the error changes (different COMport number), try again. If the error is the same, there might be an actual error with the device.
It might happen, that the BFP slider sometimes freezes and does not move the lens in position. This is because the device that controls it is entering some error state and cannot move anymore. To solve this:
You can now reopen MicroManager and resume your imaging session.
# Labels # Configuration presets # Group: Camera # Preset: Conventional ConfigGroup,Camera,Conventional,Evolve512,Gain,1 ConfigGroup,Camera,Conventional,Evolve512,MultiplierGain,1 ConfigGroup,Camera,Conventional,Evolve512,PMode,Frame Transfer Parallel Clocking ConfigGroup,Camera,Conventional,Evolve512,Port,Normal ConfigGroup,Camera,Conventional,Evolve512,ReadoutRate,5MHz 16bit # Preset: EM100 10MHz ConfigGroup,Camera,EM100 10MHz,Evolve512,Gain,1 ConfigGroup,Camera,EM100 10MHz,Evolve512,MultiplierGain,100 ConfigGroup,Camera,EM100 10MHz,Evolve512,PMode,Frame Transfer Parallel Clocking ConfigGroup,Camera,EM100 10MHz,Evolve512,Port,Multiplication Gain ConfigGroup,Camera,EM100 10MHz,Evolve512,ReadoutRate,10MHz 16bit
COMport | Baud rate | Settings | |
A405 | COM31 | 57600 | |
A488 | COM28 | 57600 | |
A561 | COM27 | 57600 | |
A638 | COM34 | 57600 | |
Mega-Servo | COM37 | 57600 | Input pull-up on |
PI | COM1 | 115200 | axis = Z, limit = 100 |
SmaractXY | COM2 | 115200 | (chan,dir)x = (1,-1), (chan,dir)y =(0,-1) |
SmaractZ | COM2 | 115200 | channel = 2, direction = 1 |
Maestro | COM23 | 57600 | (max,min,ServoNr)=(2000,990,5) |
Luxx405 | COM36 | 500000 | |
Luxx488 | COM29 | 500000 | |
Luxx638 | COM35 | 500000 | |
Cobolt561 | COM33 | 115200 | max power = 300 |
Once only:
Function | Value |
---|---|
Sensor Mech. Correction 1 in Sensor Mech. 2 | 0x02000200 = 1 |
Sensor Mech. Correction 2 in Sensor Mech. 2 | 0x02000200 = +/- 0,01 |
Soft Voltage Low Limit 1 | -30 |
Soft Voltage High Limit 1 | 130 |
Soft Voltage Low Limit 2 | -10 |
Soft Voltage High Limit 2 | 110 |
Switch to external sensor:
Switch back to internal sensor:
Prozedur: 1. Scannen des kompletten Bereiches des P-726 (100µm) in closed loop (über den internen Sensor geregelt). Dabei den externe Sensor überwachen bis dieser ein Messsignal hat. 2. Auf „Messposition“ fahren. 3. Piezospannung abfragen (VOL? 1) 4. Regler ausschalten. (die Piezospannung bleibt auf dem letzten Wert) 5. Matrix + Limits auf externen Sensor in den flüchtigen Speicher schreiben: Pos Sensor 1: 0x07000500 = „0“ ; Anpassung der Sensormatrix Pos Sensor 2: 0x07000500 = „1“
Soft Voltage Low Limit: 0x0c000000 = „[gemessene Spannung – 5V]“ ; maximal Begrenzung der Spannung für den Fehlerfall z.B. außerhalb des Sensorsignals Soft Voltage High Limit: 0x0c000001 = „[gemessene Spannung + 5V]“ 6. Sollwertvorgabe analoger Eingang: 5V = 0µm 7. Regler einschalten. (=Autofocus ein)
Achtung: Beim Umschalten auf internen Sensor, müssen die Soft Limits und die Sensor Matrix wieder auf den ursprünglichen Wert gesetzt werden.
ConfigGroup,Camera,EM200 20MHz,Evolve512,ClearMode,Pre-Sequence ConfigGroup,Camera,EM200 20MHz,Evolve512,Gain,1 ConfigGroup,Camera,EM200 20MHz,Evolve512,MultiplierGain,200 ConfigGroup,Camera,EM200 20MHz,Evolve512,PMode,Frame Transfer ConfigGroup,Camera,EM200 20MHz,Evolve512,Port,Multiplication Gain ConfigGroup,Camera,EM200 20MHz,Evolve512,ReadoutRate,20MHz 16bit
ConfigGroup,Camera,Conventional,Evolve512,Gain,1 ConfigGroup,Camera,Conventional,Evolve512,MultiplierGain,1 ConfigGroup,Camera,Conventional,Evolve512,PMode,Frame Transfer ConfigGroup,Camera,Conventional,Evolve512,Port,Normal ConfigGroup,Camera,Conventional,Evolve512,ReadoutRate,5MHz 16bit ConfigGroup,Camera,Conventional,Evolve512,ClearMode,Pre-Sequence
ConfigGroup,Camera,EM200 10MHz,Evolve512,Gain,1 ConfigGroup,Camera,EM200 10MHz,Evolve512,MultiplierGain,200 ConfigGroup,Camera,EM200 10MHz,Evolve512,PMode,Frame Transfer ConfigGroup,Camera,EM200 10MHz,Evolve512,Port,Multiplication Gain ConfigGroup,Camera,EM200 10MHz,Evolve512,ReadoutRate,10MHz 16bit ConfigGroup,Camera,EM200 10MHz,Evolve512,ClearMode,Pre-Sequence
ConfigGroup,Camera,EM100 10MHz,Evolve512,Gain,1 ConfigGroup,Camera,EM100 10MHz,Evolve512,MultiplierGain,100 ConfigGroup,Camera,EM100 10MHz,Evolve512,PMode,Frame Transfer ConfigGroup,Camera,EM100 10MHz,Evolve512,Port,Multiplication Gain ConfigGroup,Camera,EM100 10MHz,Evolve512,ReadoutRate,10MHz 16bit ConfigGroup,Camera,EM100 10MHz,Evolve512,ClearMode,Pre-Sequence
Intializing axes (necessary after rebooting controller):
Important: before moving the actuators, check that the plate is perfectly sitting on them (actuator tips in the small rails underneath the plate) !!
+49 157 56 74 58 86
languages: english, polski, a bit german, 3 words of french
For instance camera not working (no ventilation) or no light on the focus-lock laser.
Sometimes you even have to unplug and replug the permanent power socket.
Comport | Com3 | Com10 | Com9 | Com11 | Com12 |
---|---|---|---|---|---|
Device | PI | Smaract 3D | Smaract 1D | ||
Baudrate | 56700 | 115200 | 115200 | ||
Use | Objective focusing | Sample and photodiode | Two color imaging | ||
Ext dialogue | PITerminal.vi |